Sensedge Duct Technical Specifications
Wireless, Battery-Powered Model
(DU-300A/W, 301A/W)
Wired Model
(DU-200P/PW, 201P/PW)
Sensor technology
Laser particle sensor (Light scattering)
Mass concentration accuracy for PM2.5
0 to 30 μg/m³: ±3 μg/m³
30 to 1000 μg/m³: ±10 % m.v.
Mass concentration accuracy for PM10
0 to 30 μg/m³: ±3 μg/m³
30 to 1000 μg/m³: ±15 % m.v.
Mass concentration range
0 to 1,000 μg/m³
Mass concentration size range
PM2.5: 0.3 to 2.5 μm
PM10: 0.3 to 10.0 μm
Sensor output resolution
1 μg/m³
Calibration
Calibrated against standardized aerosol mix
WELL specification requirements
Adjustable particle density (K-factor) to accommodate project/region-specific particulate profile.
Complies with WELL Performance Verification Guidebook to be used in WELL certification.
Sensor technology
Multi-pixel metal oxide sensor (MOx)
Accuracy
±15 % ±4 ppb
±15 % ±18 ug/m3
Measurement range
0 - 1,382 ppb
0 - 5,482 ug/m3
Sensor output resolution
1ug/m3
Target gas profile
Complex mixture of 22 VOCs1 as defined by Molhave et al.
Sampling process
Calibration
Ethanol in clean air
WELL specification requirements
Calibration gas: ethanol
Target gas profile (ppb=µg/m³ conversion factor under STP): 22 VOC mixed per Molhav et al. (1 ppb = 4.57 µg/m³)
Complies with WELL Performance Verification Guidebook to be used in WELL certification.
Sensor technology
Non-dispersive infrared (NDIR)
Accuracy4
± 40 ppm ± 3%
Comply with ANSI/ASHRAE Standard 62.1-2022
Measurement range
400 to 5,000 ppm2
Up to 10,000 ppm extended range3
Sensor output resolution
1 ppm
Target gas
CO2
Sensor technology
Electrochemical
Accuracy
0 to 20 ppm: ±1 ppm
>20 ppm: ±5%
Measurement range
0 to 100 ppm
Sensor output resolution
0.1ppm
Target gas
CO
Sensor technology
Electrochemical
Accuracy
0 to 100 ppb: ±10 ppb
>100 ppb: ±10 %
Measurement range
0 to 2,000 ppb
Sensor output resolution
1 ppb
Target gas
O3
Sensor technology
Electrochemical
Accuracy5
0 to 100 ppb: ±10 ppb
>100 ppb: ±10 %
Measurement range
0 - 2,000 ppb
Sensor output resolution
1 ppb
Target gas
NO2
*(Available in DU-200PW, 201PW, 300W, 301W)
Sensor technology
Digital sensor
Accuracy
±0.2 °C
Measurement range
-20 - 100 °C
Sensor output resolution
0.1 °C
Sensor technology
Digital sensor
Accuracy
±2 % RH
Measurement range
0 - 100 % RH
Sensor output resolution
0.1 % RH
Sensor technology
Differential pressure (Thermal mass flow)
Filter pressure drop (2 channels)
Range: ±500Pa
Accuracy: 3%
Resolution: 0.1Pa
Duct air flow (1 channel)
Range: 1.5 - 20m/s
Accuracy: ±10%
Resolution: 0.1m/s
Wireless Model
(DU-300A/W, 301A/W)
Battery: 6 x Li/SOCl2 AA size
Battery life:
- Up to 8 years of battery life6 using Adaptive Sampling™
- 3.5 years using a high frequency fixed sampling rate7, 8
USB-C: 5V 0.5A (Cable not included)
PoE: Via PoE to USB-C converter (Cable not included)
Wired Model
(DU-200P/PW, 201P/PW)
Direct Wiring: 12 - 30 V DC
USB-C: 100~240V AC (5V 1.8A DC)
PoE:
- IEEE 802.3af (PoE), Class3
- IEEE 802.3at (PoE+), Class3
- PD maximum power ≤ 10 W
- PSEs: Midspan & endspan supported
- Cable: Cat5 (Cat5e, Cat6, and Cat6a)
Power consumption
- Average: 1.7W
- Peak: 4~5W
Wireless Model
(DU-300A/W, 301A/W)
Frequency range (MHz):
- Sub-Ghz wireless communication
- IN865/EU868/US915/AU915/KR920/AS923
- These frequency bands cover over 200 countries, contact Kaiterra for details on compliance in your location.
Gateway: supports cellular9 and ethernet
Wired Model
(DU-200P/PW, 201P/PW)
Wi-Fi
- 2.4 GHz 802.11 b/g/n
- Security supported: 64/128 WEP, WPA-PSK, WPA2-PSK, WPA, WPA2 Personal
Ethernet
- IEEE 802.3
- Data rate: Up to 100 Mbps
Wireless Model
(DU-300A/W, 301A/W)
BACnet/IP via Gateway
Open API
Wired Model
(DU-200P/PW, 201P/PW)
BACnet/IP
RS-485 Modbus/RTU
Cloud MQTT
On premise MQTT
Open API
Wireless Model
(DU-300A/W, 301A/W)
Data Storage
- Cloud storage
- Local data storage: 0.5 days
Data Sampling
- Adaptive Sampling™ to automatically adjust sampling frequency to maximize battery life.
- Selected sensors support a configurable sampling rate from 1 minute to 24 hours.
Wired Model
(DU-200P/PW, 201P/PW)
Data Storage
- Cloud storage
- Local data storage: 1 hour
Data Sampling
Frequency of readings (Log interval):
- 1 minute, 1 hour, 1 day
Data push interval:
- 1 minute10
Compatible Modules (Wireless Model)
KM-300: Particulate Matter sensor module
KM-308: CO, O3, NO2 sensor module
Compatible Modules (Wired Model)
KM-200: Particulate Matter sensor module
KM-208: CO, O3, NO2 sensor module (Coming soon)
Calibration
Swappable sensor modules
Kaiterra’s platform architecture meets the most stringent security standards and is regularly subjected to 3rd party penetration tests. Read more about our security here.
Length
375 mm (14.76 in)
Height
300 mm (11.8 in)
Width
300 mm (11.8 in)
Weight
Wireless Model:
Without filter efficiency feature:
1488 g (3.28 lbs)
With filter efficiency feature:
2370 g (5.22 lbs)
Wired Model:
Without filter efficiency feature:
1470 g (3.24 lbs)
With filter efficiency feature:
2388 g (5.26 lbs)
Duct mount with screws (on supply and/or return air ducts)
Operational temperature: -20 - 50 °C
Operational humidity: 5 to 95 %RH, non-condensing
- n-Hexane, n-Nonane, n-Decane, n-Undecane ,1-Octane, 1-Decene, Cyclohexane, m-Xylene, Ethylbenzene, 1,2,4-Trimethylbenzene, n-Propylbenzene, a-Pinene, n-Pentanal, n-Hexanal, Iso-propanol, n-Butanol, 2-Butanone, 3-Methyl-3-butanone, 4-Methyl-2-pentanone, n-Butylacetate, Ethoxyethylacetate, 1, 2-Dichloroethane
- Extended exposure to concentrations below 400 ppm may result in incorrect operation of ABC algorithm and should be avoided.
- Sensor provides readings in the extended range but the accuracy may be lower than that specified in the table.
- The accuracy specification covers environments ranging from 0-50°C and 0-80% RH, and complies with indoor air quality standards ANSI/ASHRAE Standard 62.1-2022 at 25°C.
- Sensor specifications are under controlled laboratory conditions. Field measurements may use localized ambient air quality and historical infiltration rates to enhance the accuracy and baseline readings of NO2 concentrations. This method is effective under typical indoor conditions but may not suit environments with persistently high indoor NO2 levels.
- Using Adaptive Sampling™ in a building with 15 devices, located in a typical North American city with mechanical ventilation and strong wireless signal strength to the gateway.
- During operating hours (9-5, Mon-Fri): one sample per minute for all sensors except particulate matter, and one sample every 10 minutes for particulate matter. Outside of operating hours: one sample every five minutes for all sensors except particulate matter sensor, and one sample every 60 minutes for particulate matter.
- Actual battery life may vary based on usage, environmental conditions, and other factors. Model SE-300W has a battery life expectancy of 3.5 years. Model SE-300L has a battery life expectancy of 8.5 years.
- Cellular is supported in select countries and regions.
- Customizable upon request.